Catalog Mitutoyo

NC Linear Scales ST36 Open scales with Sine & Square-Wave Output This is a high resolution, high accuracy, portable model designed for exposed installations. The ST36 offers you the following benefits: • High reliability with stable signal output. • Outputs two-phase sinusoidal signal (signal pitch 4μm) and two-phase square-wave signal simultaneously. • Outputs two-phase sinusoidal signal (signal pitch 4μm) and two-phase square-wave signal simultaneously. • You are provided with a compact interface box as standard. Installation dimensions are available from Mitutoyo – please contact. ST36 Model A/C PA PA PB PB PZ PZ t 70 10µm Model B/C Specifications Accuracy 10 - 300 mm ±0,5 μm 350 - 500 mm ±1 μm 600 - 1000 mm ±2 μm 1100 - 3000 mm ±2 μm/m Output waveform Two 90° phase-shifted sinusoidal signals 2Vpp (Model A and C) Two 90° phase-shifted square-wave signals (Model B and C) RS422 Line Driver Scale reference points 10 -75 mm Center point scale 100 - 3000 mm every 50 mm Max. drive speed 1200 mm/sec Power supply 5V DC ± 10% Signal period 8 μm Cross section 22 x 14,9 mm Resolution 0,01; 0,02; 0,05; 0,1 μm Signal period 4 μm Refer to the NC System leaflet for more details All products to be sold to commercial customers. Product illustrations are without obligation. Product descriptions, 388 All products to be sold to commercial customers. Product illustrations are without obligation. Product descriptions, in particular any and all technical specifications, are only binding when explicitly agreed upon.

RkJQdWJsaXNoZXIy MTU4ODc4MA==